TURBISCAN DNS is the first ready-to-use platform for dispersibility and stability studies. Dispersibility studies can be performed, while processing the dispersion (at flow) and on the native sample (no dilution, no added forces) via Static Multiple Light Scattering (SMLS), as recommended by the ISO TS 21357.
The flow studies can be done by either pumping the sample from the process or reactor (circulation function of the Turbiscan DNS: T-LOOP) or by mixing directly in the measurement cell (mixing function: T-MIX).
The stability measurement is done at rest and the Turbiscan DNS offers a fast (up to 1,000 times faster), quantitative, and reliable stability measurement of destabilization kinetics without dilution or mechanical forces.
PHARMACEUTICAL
CHEMISTRY / POLYMERS
FOOD & BEVERAGES
COATINGS
ELECTRONICS
PETROLEUM
For Dispersibility ‘n’ Stability study, using our unique SMLS technology
Online particle determination on native sample and without any dilution at high frequency (up to 10 measurements per second),
Accelerated destabilization detection, up to x1000 faster than visual observation,
Redispersion studies can be performed easily and repeatedly
The Online module, TLoop and TMix, provides two possibilities for online studies: either connect the DNS to your process via the Turbiscan Loop, or use vial as mini reactor by equipping it with the specific blade
Click and analyse: The instrument is fully integrated platform, everything included and everything is perfectly aligned
— LIMITLESS APPLICATION RANGE: CONCENTRATION FROM 10-4 TO 95% V/V AND PARTICLE SIZE RANGE FROM 10 NM TO 1 MM
— ONLINE MEASUREMENT: MIXING AND PUMPING FUNCTION AVAILABLE
— MEASUREMENT MODE: SCAN MODE FOR STABILITY MEASUREMENT AND FAST MODE FOR ONLINE MEASUREMENT (UP TO 10 MEASUREMENTS/PER SECOND).
— TEMPERATURE CONTROLLED FROM RT TO 60°C (FOR STABILITY AND ONLINE MEASUREMENT)
— SAMPLE STABILITY & PARTICLE SIZE ANALYSIS
— INTUITIVE SOFTWARE & TSI CALCULATION
— MEASUREMENT CELLS VOLUME: 20ML AND 4ML (ONLY FOR STABILITY) + POSSIBILITY TO DEVELOP SPECIFIC CELLS HOLDERS
Technology | SMLS 880nm |
Sample cell volume | 4ml or 20ml |
Measurement while process | Yes: Mixing & Circulation |
Temperature range | RT - 60oC |
Number of samples | 1 |
Maximum sample concentration | %95 v/v |
Measured size range | 10nm - 1mm |
Reproducibility / Repeatibility | %0.1 / %0.05 |
Acquisition scan step | 40µm |
Maximum measurement frequency | 10 measurements/s |
ISO compliance | TR 13097, TR 18811, TR 13014, TS 21357 compliant |
Dimensions | 700 x 630 x 520 mm |
Weight | 30kg |